The information contained in this manual may be confidential and proprietary and is the property of GE Analytical Instruments. Information disclosed herein shall not be used to manufacture, construct, or otherwise reproduce the goods disclosed herein. The information disclosed herein shall not be disclosed to others or made public in any manner without the express written consent of GE Analytical Instruments.
5:00 p.m. (Mountain Time), excluding all company and legal holidays. Telephone support is provided for troubleshooting and determination of parts to be shipped from GE Analytical Instruments to the customer in order to return the product to oper- ation. If telephone support is not effective, the product may be returned to GE Analytical Instruments for repair or replace- ment.
The Operation and Maintenance Manual is believed to be accurate at the time of publication and no responsibility is taken for any errors that may be present. In no event shall GE Analytical Instruments be liable for in- cidental or consequential damages in connection with or arising from the use of the manual and its accompanying related materials.
œuvre de procédures de maintenance non recommandées peut annuler toute disposition de garantie. Les procédures de correction indiquées dans le présent document sont les seuls remèdes du client. Le groupe GE Analytical Instruments ne saurait en aucun cas être tenu pour responsable de tout préjudice direct, indirect ou spécial de quelque na- ture que ce soit (y compris, les pertes de bénéfices), qu’il soit fondé...
Gli unici rimedi spettanti all’utente sono quelli qui inclusi. In nessun caso GE Analytical Instruments sarà responsabile per dan- ni diretti, indiretti, speciali, accidentali o consequenziali (inclusa la perdita di profitti) risultanti dall’applicazione del contratto, atto illecito o altra teoria legale.
Page 23
Always “clean-up” the Analyzer by running low-TOC DI water after running high TOC or salt samples. Warning If this instrument is used in a manner not specified by GE Analytical Instruments, the protection provided by the instrument may be impaired. Warning The glass window in the UV reactor indicates proper lamp operation and should not be viewed too closely (closer than 6 inches) or for extended periods (greater than 1 minute).
Page 25
Este producto es de clase de seguridad I. Debe conectarse a una fuente de alimentación con toma de masa. Advertencia Si este instrumento se utiliza de una manera no especificada por GE Analytical Instruments USA, la protección ofrecida por el instrumento puede verse reducida.
Page 27
Ce produit est de sécurité – classe I. Il doit être relié à une source d’alimentation mise à la terre. Avertissement Si cet instrument est utilisé de manière non conforme à ce qui est spécifié par le groupe GE Analytical Instruments aux USA, la protection fournie par l’instrument risque d’être réduite.
Page 29
Dies ist ein Produkt der Sicherheitsstufe I. Es muss an eine geerdete Stromquelle angeschlossen werden. Warnung Wenn dieses Instrument in einer Art und Weise verwendet wird, die nicht von GE Analytical Instruments USA festgelegt ist, kann der durch dieses Instrument gebotene Schutz beeinträchtigt werden.
Page 31
Questo è un prodotto che rientra nella classe di sicurezza I. Deve essere collegato a una sorgente di alimentazione con messa a terra. Avvertenza Se lo strumento viene utilizzato in modo diverso da quello specificato da GE Analytical Instruments USA, la protezione fornita dallo strumento può risultare compromessa. Avvertenza Interrompere sempre l’analisi prima di spegnere o scollegare l’analizzatore.
Chapter 1. Introduction The Sievers 500 RL On-Line TOC Analyzer from GE Analytical Instruments is a high-sensitivity Analyzer used to measure the concentration of total organic carbon (TOC), total inorganic carbon (TIC), and total carbon (TC = TOC + TIC) in water samples (for patent information, see the section called “Trademarks and Patents” on page 14).
Chapter 2: System Description System Overview The Sievers 500 RL On-Line TOC Analyzer consists of five major subsystems: Sample inlet system and sample pump, including the Integrated On-Line Sampling (iOS) System or Super iOS System. Oxidation reactor Measurement module, comprising: •...
Analyzer via the USB port to a USB flash storage device (provided) and then transferred to any computer that supports USB. The Sievers 500 RL On-Line TOC Analyzer also has three analog outputs (4-20 mA) which can be customized to track specific data values, four alarms, and two binary ports.
Chapter 3. Installation Overview This chapter provides installation instructions for the Sievers 500 RL On-Line TOC Analyzer and will help familiarize you with Analyzer design and function. An interior overview diagram is available in the “Maintenance” chapter on page 140. If you need additional assistance, contact GE Analytical Instruments Technical Support at 800.255.6964.
Analyzer from the box. Next, verify that all items are present: Sievers 500 RL On-Line TOC Analyzer Operation and Maintenance Manual Validation Support Package (VSP) Volume I...
Step 3: Select a Location for the Analyzer The Sievers 500 RL On-Line TOC Analyzer is designed to be mounted on a wall or support stand. Allow a minimum of 5 cm clearance between the back of the Analyzer and the wall for heat dissipation; allow 30.5 cm of clearance on the sides, top, and bottom of the analyzer for the plumbing and electrical connections.
Analyzer. Connecting to a Power Supply Installation of the Sievers 500 RL On-Line TOC Analyzer requires an external source of AC power connected to the enclosure using a water-tight conduit connector. The electrical connection should be performed by a qualified electrician.
Installing the Printer The Analyzer has a 25-pin parallel (Centronics) printer port for connecting an optional printer. The printer must be IBM compatible or Epson FX 850 compatible. GE Analytical Instruments sells Epson, Seiko, and Citizen printers for use with the Analyzer.
Enabling DataGuard or Password Protection (Optional) The Sievers 500 RL On-Line TOC Analyzer offers two levels of security, one included with the Analyzer and the other available for purchase as an upgrade from GE Analytical Instruments. The Password Protection feature is included with the Analyzer and provides a basic level of security.
Chapter 4. Basic Analyzer Operation Overview The Sievers 500 RL On-Line TOC Analyzer utilizes a touch-sensitive color LCD for all menu selection activities. An overview of the menu structure is given in “The Menu Screen” on page 69. An overview of hardware features starts on page 96.
Data area Status area Figure 10: The Sievers 500 RL On-Line TOC Analyzer Main Screen The Main screen is divided into three areas: The Header contains the name of the screen, the date and time, and status icons representing the following conditions: ...
Setting Up Analog Output The Sievers 500 RL On-Line TOC Analyzer has three 4-20 mA outputs. Select the output range for the 4-20 mA analog outputs by following the steps below. Instructions for wiring the 4-20mA output can be found in the “Installation”...
Analyzer, see “Using HyperTerminal” on page 179 in the Appendix. Note: For long distance connections (up to 1,000 feet or 305 meters), GE Analytical Instruments recommends use of either a RS-232 current loop converter or a RS-422 converter. You can also use an Ethernet connection.
Chapter 4: Basic Analyzer Operation Displaying System Information If you need to contact GE Analytical Instruments about your Analyzer, you can find most essential information on a single screen. Select the Maintenance tab. Press the System Info button. The following unit information is displayed: •...
TOC values, or an increase in the standard deviation values. The iOS System GE Analytical Instrument offers three inlet systems for use with the Sievers 500 RL On-Line TOC Analyzer: a sample inlet block for dedicated on-line analysis, a standard Integrated On-Line Sampling (iOS) System, with one vial port, and a Super iOS, with four vial ports.
Chapter 5. Password Protection and DataGuard Overview The Sievers 500 RL On-Line TOC Analyzer includes Password Protection using one user name and password for all users. You can activate or deactivate this feature, as needed. Using Password security, there are two options for Menu Lockout (On and Off) that define when a user must enter the user name and password.
DataGuard offers support for regulation 21 CFR Part 11, providing multiple levels of user access to the Analyzer via up to 100 unique User IDs and maintaining an audit trail of all user operations. DataGuard must be purchased as an upgrade from GE Analytical Instruments. Once it is enabled, it cannot be disabled. Enabling DataGuard DataGuard is locked until it is activated by following the steps below.
Note: the Analyzer. If you forget the login information for the new administrator account that you have created, contact GE Analytical Instruments for assistance. Adding User IDs After DataGuard has been enabled, you can add a User ID by following these steps: Select the DataGuard tab.
The Sievers 500 RL On-Line TOC Analyzer can be calibrated using one of two methods: single-point calibration at 1.5 ppm; or multi-point calibration at 500 ppb, 1.0 ppm, and 1.5 ppm. GE Analytical Instruments recommends single-point calibration for most applications.
(System Suitability, Accuracy & Precision, Linearity, Robustness, Specificity, and JP Protocol) can be purchased individually, or as a combined Validation Set. To purchase standards in North America or Canada, contact GE Analytical Instruments by phone at 800.255.6964 or 303.444.2009. To purchase standards in the United Kingdom, contact GE Analytical Instruments at 44(0) 161866 9337.
Warning Standards for the Sievers 500 RL On-Line TOC Analyzer are acidified. If you prepare your own standards and use glass vials, ensure that you acidify the standards to a pH of 3, or the system protocols will yield poor results.
Chapter 6: Calibration and System Protocols Accuracy, Precision, and Calibration Verification The Sievers 500 RL On-Line TOC Analyzer automatically calculates accuracy, precision, and calibration verification values. Use this procedure to confirm that the Analyzer’s current calibration is accurate. GE Analytical Instruments recommends performing a verification after replacement of consumables, such as the pump heads (with pump tubing), DI water cartridge, or UV lamp, and after calibration.
Chapter 6: Calibration and System Protocols Performing Pharmacopeia Tests The Sievers 500 RL TOC Analyzer is designed to comply with various Pharmacopeia testing. Additional Method Specificity and Method Robustness protocols, compliant with ICH Q2(R1), are also available as optional protocols from GE Analytical Instruments.
Refer to Table 13 below for the recommended maintenance schedule. The in-line filter, UV lamp, DI water cartridge, and sample pump tubing (including pump heads) must be purchased from GE Analytical Instruments. If you need additional assistance, contact GE Analytical Instruments Technical Support at 800.255.6964. Training is also available from GE Analytical Instruments.
The sample pump tubing and pump heads must be purchased from GE Analytical Instruments; use of tubing from other sources or failure to replace the tubing on the prescribed replacement schedule will affect Analyzer functionality.
If the filter element clogs too frequently, contact GE Analytical Instruments to receive help in the selection of larger-capacity filters. As the filter is used, the flow rate of water through the iOS System will decrease and can even stop.
This chapter provides the starting point for troubleshooting basic issues with your Analyzer. If you need additional assistance, in the U.S. contact GE Analytical Instruments Technical Support at 800.255.6964. In other parts of the world, contact your local GE Analytical instruments representative.
Explanation: Based on the history of usage, the Analyzer estimates that only 10% of the resin bed life remains. Order a new resin bed from GE Analytical Instruments now to avoid down time when the resin bed expires. 8 - User List Warning Message: “The user list is nearly full.
Page 155
25 - History Data Warning Message: “A problem has been corrected in the history data. Contact GE if the warning persists.” Explanation: The Analyzer has detected a problem reading or writing to the data history memory; as much of the data history as possible is reconstructed.
Page 156
Analyzer’s accessories kit, it is possible your USB device is not compatible with the Analyzer. For a list of known compatible USB flash memory drives, follow the Sievers 500 RL Analyzer link on our Web page, www.geinstruments.com.
Page 157
System in the correct order. Consult Chapter 6: Calibration and System Protocols for the correct vial order. 56 - Sample Pump Error Message: “The sample pump is not pumping. Please refer to the Oper. and Maint. Manual. Contact GE is problem persists.”...
Page 158
Autozero when two consecutive negative measurements occur. If negative readings occur after a second TOC Autozero, this warning is issued, and the Analyzer will abort analysis. Consult troubleshooting techniques under “Step 3: Review Solutions for Basic Problems” on page 159, and contact GE Analytical Instruments Technical Support if negative readings persist.
Note that if tubing has become bent, kinked, or otherwise damaged, a tubing replacement kit is available from GE Analytical Instruments Technical Support. Step 3: Review Solutions for Basic Problems If a visual inspection of the Analyzer does not help you identify the source of the problem you are experiencing, consult the specific problems discussed in this section.
While the Analyzer is operating, open the Analyzer front panel and make sure the sample pump is turning. If the pump is not turning and the Analyzer is operating (the display screen indicates that analysis is taking place), contact GE Analytical Instruments Technical Support for instructions. Warning To avoid potentially dangerous shock, do not touch anything inside the Analyzer while observing the sample pump.
Press the Test Pumps button, and then press the DI Pump button. Cycle the pump on and off a few times, by pressing the On and Off buttons, finally leaving it set to On. If there is no flow, contact GE Analytical instruments for further assistance.
In some instances, after consulting with GE Analytical Instruments Technical Support, it will be necessary to return the Analyzer to the factory for repairs. Only return the Analyzer to GE Analytical Instruments if Technical Support has issued you a Return Authorization (RA) number.
Page 168
22. Close and latch the Analyzer front panel. You can now repackage the Analyzer and ship it to GE Analytical Instruments. For international shipments, coordinate with an GE Analytical Instruments representative to ensure quick passage through customs.