Certificate of Quality HYC-610 HYC-890 Checker: HYC-940 HYC-940F Haier Medical and Laboratory Products Co., Ltd. Address: Haier Industrial Park, Economic Technology Development Zone. Qingdao 266510.P.R.China Web:www.haiermedical.com Read the Operation Manual carefully before using your appliance. Version:1st,2017 Keep the Operation Manual in a safe place.
HYC-890 HYC-940 Shelves HYC-940F Label strip Manufactures Name : Haier Medical and Laboratory Products Co.,Ltd. Manufactures Address : Haier Industrial Park, Economic Technology Drain hole cap Development Zone. Qingdao 266510.P.R.China Fixed card Refer to in this declaration confirms with the following directive(s)/standards:...
Page 3
User-friendly design, computer control, smart and carefree, adjustment User-friendly not required. Design High performance thermal insulation. Electric heated glass door with multiple anti-condensation technologies(except HYC-260/360). Automatic removal of condensate water. Note: Technical information might be somewhat different on your refrigerator than published due to continuous improvement.
Safety Precautions Dear users: Specification Thanks for your choice of Haier pharmaceutical refrigerator, please make sure you have carefully read and observed the contents with following signs in the manual, for better understanding of this Name Pharmaceutical Refrigerator manual and better use of this product, so as to prevent personal injuries and refrigerator damage.
Page 6
Specification Do not place containers of water or heavy objects on the refrigerator. Falling objects may cause personal injury and overflown water may damage the insulation to cause electric shock and fire. Name Pharmaceutical Refrigerator Do not ground the refrigerator through gas pipes, water supply pipes, telephone lines or Model HYC-610 lighting conductors.
Page 7
Specification Name Pharmaceutical Refrigerator Always check the settings in the controller after restarting a unit from a power outage or shut off. Change of settings may cause damage to the products stored. HYC-390 Model HYC-290 HYC-390F Exterior Dimensions If the power is shut off, let the unit sit for 5 minutes before it is powered up again to avoid 665×710×1965 665×710×1965 665×710×1665...
The refrigerator is designed to operate at a condition of 16 ℃ to 32 ℃ and humidity less than Interior light Fluorescent lamp 15W 85%RH (for HYC-260/360,the humidity should be less than 70%RH). Small amount of condensation Net Weight 88kg...
Ambient temperature: 16 ℃ to 32 ℃, the range of 18 ℃ to 25 ℃ is optimal. Air conditioning system is required as necessary. Ambient humidity: below 85%RH(for HYC-260/360,the humidity should be less than 70%RH). Avoid excessive dust. Avoid mechanical vibration.
Preparation before usage Circuit diagram(HYC-890) 1. Remove all packaging materials. For safety reason, the equipment of HYC-890/940/940F is situated on a wooden pallet and secured by metal brackets. Please remove the screws and place the metal brackets under the unit.(Pic.1). Forklift or specialty lifting equipment should be Pic.1 used to remove the unit off the pallet.
Page 11
6. Installation of Shelves and Label strip HYC-260/360/610/890/940/940F Take out the shelf, label strip and rack fastener from the refrigerator. Fix the shelf fasteners on the liner plate at the appropriate distance and height.
Page 12
8.Back Bracket(HYC-260/360/290/390/390F) Circuit diagram(HYC-290/390/390F) HYC-260/360 Install the support bolts into the mounting holes on the back of the refrigerator. (See Pic.3) Pic.3 HYC-290/390/390F Remove four M5 bolts from the back of cabinet,then fix the back bracket for rear wall distance with the removed bolts.(See Pic.4)...
Circuit Diagram Initial Power-on Circuit diagram(HYC-260/360) Follow the flow chart below to start the refrigerator. While the refrigerator is empty, connect the power line to the appropriate dedicated socket and make sure the power supply in the socket matches the required voltage and frequency.
Page 14
If you have any questions about the operation of the refrigerator, please refer should be checked and recorded using appropriate methods. When the refrigerator to the table frequent asked questions as follows. Call Haier technical support temperature is out of specification, take proper actions to protect the stored samples.
You can purchase a high quality battery from a store for the same size and voltage(9V). You may HYC-260/360 also call Haier to purchase a high quality replacement battery. The lead time to deliver the battery is Light switch 15 days after receiving the payment. Please call Haier for details.
When you are running short of the chart paper, please contact Haier to purchase more. The lead time to deliver the charts is 15 days after receiving the payment . Please call Haier for details.
Page 17
If the alarm condition still exists, the buzzer will return to the alarm mode again USB port automatically 20 minutes later. HYC-260/360 series does not have the remote alarm function. Battery switch Remote Alarm Terminals HYC-290/390/390F/610/890/940/940F has remote alarm terminals:...
Page 18
Alarm Alarm or Safety Phenomenon Alarm Indication Buzzing Alarm HYC-610 If upper temperature sensor or lower High temperature sensor perceives Alarm light flashes Impulsive sound alarm temperature temperature of the refrigerator ≥8℃ Light switch Control panel If upper temperature sensor or lower temperature sensor perceives Alarm light flashes Impulsive sound alarm...
Page 19
HYC-890 1. If control buttons are not touched in 60 seconds, all changes of parameters will not become effective. The previous settings are restored. Light switch USB port 2. If the unit is powered off before the parameters are saved, the control parameters in Door lock Control panel the RAM will be lost.
Page 20
shows“1P” once again. Users can set 1P-5P again. After the setting, press the “Cal ” for 5s, and all s HYC-940 ettings will be automatically saved. And then the display area of inner temperature shows the inner temperature normally. Light switch Note: during the said 1P-5P setting process, parameters can be specifically regulated via the Door lock Control panel...
Page 21
HYC-940F USB Function Functions of USB interface The computer has a USB function. Test data can be exported via the USB port. The computer can Light switch Control panel automatically collect and store the test data within the recent 10 years. It collects data every other 6 minutes and stores them automatically.
If there is a need to be adjusted, perform the following procedures. 1. Let the unit run for at least 2 additional hours after it reaches stable condition, HYC-260/360/290/390/390F/890/940/940F 2. Place a standard thermocouple in the temperature solution bottle filled with 100 ml 10% glycerol solution.
Application method Maintenance and replacement of the recorder's standby battery Temperature display When the recorder’s LED light blinks green, check the recorder performance as follows. 1) Remove the standby battery. If the LED light is off and the recorder stops working, the This refrigerator has been set capable of reaching appropriate temperature (2~8℃) automatically main power supply of the recorder fails.
Page 24
(AC) is switched on and its stand-by power 2) Forced defrost cycle(HYC-260/360/290/390/390F only have this funtion) battery is replaced. When the standby battery is low, LED indicator light of the recorder blinks to When the environmental humidity is too high or the products loaded into the refrigerator emit much indicate that the battery shall be replaced.
Need help?
Do you have a question about the HYC-260 and is the answer not in the manual?
Questions and answers